SUBJECT: Emergency
Use of an Unapproved Medical Device
An
unapproved medical device may be used in human subjects only if it is approved
for clinical testing under an approved application for an Investigational Device
Exemption (IDE) under section 520g of the
Federal Food, Drug, and Cosmetic Act [21USC360(j,g) and 21CFRpart812; 45CFR]. Medical
devices that have not received marketing clearance under section 510k of the
FD&C Act are also considered unapproved devices that require an IDE. The Food and Drug Administration (FDA)
recognizes that emergencies arise where an unapproved device may offer the only
possible life-saving alternative, but an IDE for the device does not exist, or
the proposed use is not approved under an existing IDE, or the physician or
institution is not approved under the IDE. Using its enforcement discretion, FDA
has not objected if a physician chooses to use an unapproved device in such an
emergency, provided that the physician later justifies to FDA that an emergency
actually existed.
Unapproved Medical Device. An
unapproved medical device is defined as a device that is used for a purpose or
condition for which the device requires, but does not have, an approved
application for premarket approval under section 515 of the Federal Food, Drug,
and Cosmetic Act [21USC360(e)].
Requirements for Emergency Use. Each of the
following conditions must exist to justify emergency use:
1. The
patient is in a life-threatening condition that needs immediate treatment;
2. No
generally acceptable alternative for treating the patient is available; and
3. Because
of the immediate need to use the device, there is no time to use existing
procedures to get FDA approval for the use.
FDA
expects the physician to determine whether these criteria have been met, to
assess the potential for benefits from the unapproved use of the device, and to
have substantial reason to believe that benefits will exist. The physician may not conclude that an "emergency" exists
in advance of the time when treatment may be needed based solely on the
expectation that IDE approval procedures may require more time than is
available. Physicians should be aware that FDA expects them to exercise
reasonable foresight with respect to potential emergencies and to make
appropriate arrangements under the IDE procedures far enough in advance to avoid
creating a situation in which such arrangements are impracticable.
In the
event that a device is to be used in circumstances meeting the criteria listed
above, the device developer should notify the Center for Devices and
Radiological Health (CDRH), Program Operation Staff by telephone (301-594-1190)
immediately after shipment is made. [Note: an unapproved device may not be
shipped in anticipation of an emergency.] Nights and weekends, contact the
Division of Emergency and Epidemiological Operations (202-857-8400).
FDA expects the physician to
follow as many subject protection procedures as possible. These include:
1.
Obtaining an independent assessment by an uninvolved physician;
2.
Obtaining informed consent from the patient or a legal
representative
3.
Notifying institutional officials as specified by institutional policies;
4.
Notifying the Institutional Review Board (HRAC); and
5.
Obtaining authorization from the IDE holder, if an approved IDE for the
device exists.
After-use Procedures. After an
unapproved device is used in an emergency, the physician should:
1. Report
to the HRAC within five days [21 CFR56.104(c); 45 CFR] and otherwise comply with provisions of the HRAC
regulations (21 CFRpart 56; 45 CFR);
2. Evaluate
the likelihood of a similar need for the device occurring again, and if future
use is likely, immediately initiate efforts to obtain HRAC approval and an
approved IDE for the device's subsequent use; and
3. If
an IDE for the use does exist, notify the sponsor of the emergency use, or if an
IDE does not exist, notify FDA of the emergency use (CDRH Program Operation
Staff 301-594-1190) and provide FDA with a written summary of the conditions
constituting the emergency, subject protection measures, and results.
Subsequent emergency use of the device may not occur unless
the physician or another person obtains approval of an IDE for the device and
its use. If an IDE application for
subsequent use has been filed with FDA and FDA disapproves the IDE application,
the device may not be used even if the circumstances constituting an emergency
exist. Developers of devices that could be used in emergencies should anticipate
the likelihood of emergency use and should obtain an approved IDE for such uses.
Consent. Even for an emergency use, the
investigator is required to obtain informed consent of the subject or the
subject's legally authorized representative unless both the investigator and a
physician who is not otherwise participating in the clinical investigation
certify in writing all of the requirements of exception for consent in emergency
research detailed in section 15.2 of this manual.
If, in
the investigator's opinion, immediate use of the test article (unapproved
medical device) is required to preserve the subject's life, and if time is not
sufficient to obtain an independent physician's determination that the four
conditions above apply, the clinical investigator should make the determination
and, within 5 working days after the use of the article, have the determination
reviewed and evaluated in writing by a physician who is not participating in the
clinical investigation. The investigator must notify the HRAC within 5 working
days after the use of the test article [21CFR50.23(c); 45CFR].