FacsAria

The FACSAria has 3 lasers and is able to detect up to 9 colors at one time as well as sort up to 4 populations at once. The FACS Aria is used for analysis of more than 4 colors and cell sorting, its primary function being cell sorting. The FACS Aria is also used when a fluorochrome is required that cannot be detected by the FACS Calibur.
Lasers
488-nm Coherent® SapphireTM solid state
633-nm JDS UniphaseTM HeNe air-cooled
407-n m Point Source Violet solid state
All lasers are Class IIIb.
Excitation Laser |
Detection Filter |
Detector |
Example |
488 nm (blue) |
488/10 (483-493 nm) |
FSC Diode |
FSC only |
|
780/60 (750-810 nm) |
PMT A |
PE-Cy7 |
|
695/40 (675-715 nm) |
PMT B |
PerCP Cy5.5 |
|
610/20 (600-620 nm) |
PMT C |
PE-Texas Red |
|
576/26 (563-589 nm) |
PMT D |
PE |
|
530/30 (515-545 nm) |
PMT E |
FITC |
|
488/10 (483-493 nm) |
PMT F |
SSC only |
|
|
|
|
633 nm (red) |
780/60 (750-810 nm) |
PMT A |
APC-Cy7 |
|
660/20 (650-670 nm) |
PMT B |
APC |
|
|
|
|
407 nm (violet) |
530/30 (515-545 nm) |
PMT A |
Alexa-Fluor 430 |
|
450/40 (430-470 nm) |
PMT B |
DAPI |
Analysis
Analysis and acquisition on the FACS Aria is done in the FACS Diva software. Using FACS Diva, data may be analyzed on dot plots and histograms with various gates.
Data Delivery
Please bring a USB drive or blank CD to save your data files. The Core Facility does not store or archive data for longer than about 6 months so it is your responsibility to get the data files and save them for future use.